Solving Large Problems With Small Biofactories

Microfluidic chemostat and turbidostat with flow rate, oxygen, and temperature control for dynamic continuous culture.

Title

Microfluidic chemostat and turbidostat with flow rate, oxygen, and temperature control for dynamic continuous culture.

Publication Type
Journal Article
Year of Publication
2011
Journal
Lab Chip
Volume
11
Issue
10
Pagination
1730-9
Date Published
2011 May 21
ISSN
1473-0189
Abstract

This work reports on an instrument capable of supporting automated microscale continuous culture experiments. The instrument consists of a plastic-PDMS device capable of continuous flow without volume drift or evaporation. We apply direct computer controlled machining and chemical bonding fabrication for production of fluidic devices with a 1 mL working volume, high oxygen transfer rate (k(L)a≈0.025 s(-1)), fast mixing (2 s), accurate flow control (±18 nL), and closed loop control over temperature, cell density, dissolved oxygen, and pH. Integrated peristaltic pumps and valves provide control over input concentrations and allow the system to perform different types of cell culture on a single device, such as batch, chemostat, and turbidostat continuous cultures. Continuous cultures are demonstrated without contamination for 3 weeks in a single device and both steady state and dynamically controlled conditions are possible.

Alternate Journal
Lab Chip
Citation Key
241

PubMed ID

21445442

Grant List