Comparison of the efficacy of steam sterilization indicators.

TitleComparison of the efficacy of steam sterilization indicators.
Publication TypeJournal Article
Year of Publication1979
AuthorsLee, CH, Montville, TJ, Sinskey, AJ
JournalAppl Environ Microbiol
Volume37
Issue6
Pagination1113-7
Date Published1979 Jun
ISSN0099-2240
KeywordsGeobacillus stearothermophilus, Indicators and Reagents, Spores, Bacterial, Sterilization, Temperature, Time Factors
Abstract

Twenty-one commercially available chemical steam sterilization indicators were processed in an empty autoclave for various times at temperatures between 240 and 270 degrees F (ca. 116 and 132 degrees C). The time required to reach a sterilized reading at each temperature was plotted on a semilogarithmic time-temperature plot and compared with the time-temperature sterilization curve for Bacillus stearothermophilus. Five of the indicators had time-temperature kinetics similar to those of B. stearothermophilus, but three of these overestimated the effect of processing. Two of the indicators overestimated the effect of processing and were less sensitive to temperature changes when was B. stearothermophilus. Thirteen of the indicators had time-temperature curves that crossed the B. stearothermophilus plot. One indicator produced such ambiguous results that no determinations could be made with it. Out of 21 indicators tested, only 2 appear to be capable of accurately integrating the time-temperature effect at temperatures between 240 and 270 degrees F. The other indicators should be used only after careful analysis of their suitability for use at a given temperature.

DOI10.1128/aem.37.6.1113-1117.1979
Alternate JournalAppl Environ Microbiol
Citation Key236
PubMed ID485144
PubMed Central IDPMC243363